Search this Site

 

About Med
Training and Demo Facility
Contract
Preclinical Research
Supporting the
Scientific Community
Meetings
MedLines Archive
Contact Us
Recent Publications
by Med Scientific Staff
  Current Med Citations
New Products
Upcoming Products
Sales
International
Sales Representatives
Medstate
Notation™ Repository
Researcher's Round Table
Jobs
Policies and Procedures


Safety Syringe Pump


 

PHM-108
Single Speed Safety Syringe Pump

Standard Speed 3.33 RPM  ::  Protection from Output and Computer Failure

The Safety Syringe Pump, PHM-108, has been provided as an alternate to the PHM-100 Syringe Pump to provide extra safety from an overdose. The input circuit allows the user to select the maximum dose time from 1 to 99 seconds. In addition the output circuit has two redundant on control circuits to compensate for a failure. The standard rate is 3.33 RPM, but other motors are available and can be changed by the user.

The PHM-108 pump is activated by a MED-PC ® 28 V output and turned on as long as the output is on. Thus the time on determines the dose. A safety stop switch disconnects power from the motor when the end of the syringe has been reached.

     Features  
     >> Protection from Output Failure
     >> Protection from Computer Failure
      Safety     Protect from overdose due to computer or power failure
      Syringes     Plastic or glass, 50ml to microliter
      Construction     Stainless steel & aluminum frame
 
    Flow Rate
    .017 ml/hr to 24.122 ml/min
 
    Amluracy
    0.50%
 
    Repeatability
    0.50%
 
    Size
    6”H X 6 ”W x 10 ”L (15.24cm x 15.24cm x 25.4cm)


 

 

 

 

 

Infusion Rates
The PHM-108 can amlommodate plastic or glass syringes ranging from a 50ml syringe to a microliter syringe. Listed below are the stock motors and their corresponding flow rates for a Multifit syringe. Other motor speeds may be specially ordered. To determine flow rates with other syringes, a calibration sheet is included with every pump.

 

 

Syringe Sizes
 

MOTOR

RPM

1ml

2ml

5ml

10ml

20ml

30ml

50ml

 

calibrated in ml/min.

 

PHM-108-20

20.00

.688

2.446

4.236

6.612

11.789

15.814

24.122

 

PHM-108-15

15.00

.516

1.835

3.177

4.959

8.843

11.861

18.092

 

PHM-108-10

10.00

.344

1.223

2.118

3.306

5.895

7.907

12.061

 

PHM-108-5

5.00

.172

.612

1.059

1.653

2.947

3.954

6.031

 

PHM-108-3.33

3.33

.115

.408

.706

1.102

1.965

2.636

4.020

 

PHM-108-2

2.00

.069

.245

.424

.661

1.179

1.581

2.412

 

PHM-108-1

1.00

.034

.122

.212

.331

.589

.791

1.206

 

calibrated in ml/hr.

 

PHM-108-1

1.000

2.063

7.339

12.708

19.854

35.369

47.443

72.367

 

PHM-108-.5

.500

1.032

3.669

6.354

9.918

17.684

23.722

36.183

 

PHM-108-.33

.33

.688

2.446

4.236

6.612

11.790

15.814

24.122

 

PHM-108-.2

.200

.413

1.468

2.542

3.967

7.074

9.489

14.473

 

PHM-108-.1

.100

.206

.734

1.271

1.983

3.537

4.744

7.237

 

PHM-108.-083

.083

.172

.612

1.058

1.652

2.948

3.954

6.030

 

PHM-108-.05

.050

.103

.367

.635

.992

1.768

2.372

3.618

 

PHM-108-.042

.042

.086

.306

.529

.826

1.474

1.977

3.015

 

PHM-108-.033

.033

.069

.245

.424

.661

1.179

1.581

2.412

 

PHM-108-.022

.022

.046

.163

.282

.441

.786

1.054

1.608

 

PHM-108-.008

.008

.017

.061

.106

.165

.295

.395

.603

                   

 

< back to top>

 


 



 

Site Updated: July 19, 2010  
 © Copyright Med Associates Inc., 1998-2010. All Rights Reserved
www.med-associates.com