(9 Speeds)
The delivery mechanism of this pump is similar to the PHM-100 with the difference being that a single speed is replaced by a nine speed motor. In addition the infusion time is selected by switches. This has the benefit of overdose safety protection since the internal circuit limits the infusion time to the preset value. The speed selections for this unit are 0.50, 1.00, 1.50, 2.00, 3.33, 5.00, 10.00, 15.00, 20.00 RPM providing up to 63 infusion rates when used with seven syringe sizes ranging from 1 ml to 50 ml. The infusion times are 0.1 to 99.9 seconds in 0.1 second increments. Thumb-wheel switches are used to set the infusion rate and time. The pump is activated by a standard MED-PC ® 28 volt output, or it can be manually operated.
| |
|
Syringe Sizes |
|
|
|
|
|
|
|
|
|
| |
RPM |
1ml |
2ml |
5ml |
10ml |
20ml |
30ml |
50ml |
|
|
|
|
|
|
|
|
|
| |
calibrated in ml/min. |
|
|
|
|
|
|
|
|
|
| |
20.00 |
.688 |
2.446 |
4.236 |
6.612 |
11.789 |
15.814 |
24.122 |
| |
15.00 |
.516 |
1.835 |
3.177 |
4.959 |
8.843 |
11.861 |
18.092 |
| |
10.00 |
.344 |
1.223 |
2.118 |
3.306 |
5.895 |
7.907 |
12.061 |
| |
5.00 |
.172 |
.612 |
1.059 |
1.653 |
2.947 |
3.954 |
6.031 |
| |
3.33 |
.115 |
.408 |
.706 |
1.102 |
1.965 |
2.636 |
4.020 |
| |
2.00 |
.069 |
.245 |
.424 |
.661 |
1.179 |
1.581 |
2.412 |
| |
1.50 |
.052 |
.184 |
.318 |
.496 |
.884 |
1.186 |
1.809 |
| |
1.00 |
.034 |
.122 |
.212 |
.331 |
.589 |
.791 |
1.206 |
| |
0.50 |
.017 |
.61 |
.106 |
.165 |
.295 |
.395 |
.603 |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
< back to top> |